International Journal on Magnetic Particle Imaging IJMPI
Vol. 12 No. 1 Suppl 1 (2026): Int J Mag Part Imag
https://doi.org/10.18416/IJMPI.2026.2603006
Proceedings Articles, ID 1016
Spherical harmonic-based receive coil array for a head MPI system
Main Article Content
Copyright (c) 2026 Onurcan Topcu, Egor Kretov , Justin Ackers , Patricia Römer, Niklas Hackelberg, Haozhe Wang, Florian Thieben, Takashi Yoshida, Mandy Ahlborg, Tobias Knopp, Matthias Graeser

This work is licensed under a Creative Commons Attribution 4.0 International License.
Abstract
A gradiometric receive coil is practically the modern standard in MPI. Although such a structure is one of the reliable and most effective ways of isolating the receive circuits from the influence of drive fields, it is challenging to implement it for the spatial sensitivity directions orthogonal to the cylinder axis of an MPI scanner. In this work, we demonstrate a concept of matrix coils based on spherical harmonic spatial sensitivity profiles. Our approach permits us to expand the number of independent receive channels in our head-sized MPI system to seven (and potentially more), while preserving important gradiometric characteristics of each channel.
Article Details
References
M. Kaul, K. M. Krishnan, H. Ittrich, G. Adam, and T. M. Buzug.
Towards picogram detection of superparamagnetic iron-oxide particles
using a gradiometric receive coil. Scientific Reports, 7(1), 2017,
doi:10.1038/s41598-017-06992-5.
[2] T. Knopp, T. F. Sattel, S.Biederer, J.Weizenecker,B.Gleich, J.Borgert,
and T. M. Buzug, Receive coil array for magnetic particle imaging, in
2011 IEEE International Symposium on Biomedical Imaging: From
Nano toMacro, 1666–1669, 2011. doi:10.1109/ISBI.2011.5872724.
[3] C. Juchem, T. W. Nixon, S. McIntyre, D. L. Rothman, and R. A. d.
Graaf. Magnetic field modeling with a set of individual localized
coils. Journal of Magnetic Resonance, 204(2):281–289, 2010,
doi:10.1016/j.jmr.2010.03.008.
[4] F. Thieben, F. Foerger, F. Mohn, N. Hackelberg, M. Boberg, J.-P.
Scheel, M.Möddel, M. Graeser, and T. Knopp. System characterization
of a human-sized 3d real-time magnetic particle imaging
scanner for cerebral applications. Communications Engineering,
3(1), 2024, doi:10.1038/s44172-024-00192-6.
[5] F. Thieben, E. Kretov, O. Topcu, H. Wang, T. Yoshida, M. Graeser,
M. Ahlborg, and T. Knopp. Normalized quantitative assessment
of receive-coil sensitivity using MPI transfer function measurements.
International Journal onMagnetic Particle Imaging IJMPI,
XX(X):XXX–XXX, 2026, doi:DOI_here.